Sample Filtration
Application Type
  SPECIAL
Application ID
  ITSP
Description
  Sample Filtration with ITSP System of Microliter Inc.
 ↳ leaptec.com  ↳ leapwiki.com

Contents

Overview

ITSP (Instrument Top Sample Prep) for Serial Solid Phase Extraction (SPE) or Sample Filtration uses a standard PAL syringe as the solvent reservoir for Microliter's disposable SPE cartridges. The combination replaces expensive dedicated SPE liquid handlers and cumbersome or inconsistent vacuum manifolds. Pressurizing multiple wells is eliminated, no loss of samples or inconsistency of extraction due to leaks or loss of vacuum. Furthermore standard SPE liquid handlers require a manual transfer of the extracted samples to a sample injector. Since the PAL itself is a sample injector this can be avoided using the automated ITSP – PAL combination.

Image:ITSP.png Image:ITSP_on_PAL.png

Significant Markets

Any labs who are doing High Performance Liquid Chromatography (HPLC) or more significantly, Ultra High Performance Liquid Chromatography (UHPLC) analysis. It is imperative that particles are eliminated to avoid blockage of HPLC lines and columns.

How it works by Microliter

ITSP (Instrument Top Sample Prep) is a consumable device that allows for a common autosampler used on most chromatographs to prepare client samples using Solid Phase Extraction (SPE) or filtration. ITSP’s septum seals and grips the needle creating a closed environment above the bedmass. By removing the volume above the column bedmass and replacing it with the analytical instrument’s syringe, solutions can be passed across the bedmass using the hydraulic pressure of the syringe plunger. The gripping action of the septum also allows for the ITSP device to be transported to any location on the autosampler. By defining waste and elution locations the device can be prepped in one location and moved to elute the collected compounds in another. Once eluted into a clean receptacle the instrument can discard the device and inject the eluate onto the analytical instrument.

The following diagrams illustrate the ITSP device and the typical set-up on an LC/MS/MS autosampler.

Image:ITSP_Components.png

Image:ITSP_Autosampler_Setup.png

Image:ITSP_Method_Development_Tray.png

Brochure

ITSP Sample Prep Automation Guide

ITSP Setup Guide


Video

Movies of ITSP on PAL in action.
Video from Microliter on ITSP with Quicktime

Image:ITSP1.png


Other Reference Material

2010 Method Development Poster ALA

Example Application Notes

Determination of Cocaine and Benzoylecgonine in Human Urine Using Automated ITSP Solid Phase Extraction and Liquid Chromatography Mass Spectrometric Detection
ITSP Extraction of Cortisol from Plasma
Autosampler-Based Sample Preparation for Just-in-Time Extraction of Tacrolimus, Sirolimus and Cyclosporine for LC/MS Analysis
AMHSA-5_Drugs_of_Abuse_ITSP_Application_Notes
Automated SPE-LC/MS/MS Assay for 25-OH Vitamin D Metabolites from Serum
Determination of Vitamin D Metabolites in Human Serum Using Automated ITSP Solid Phase Extraction and Liquid Chromatography-Tandem Mass Spectrometry
Development of a SPE/DART Assay for Rapid Narcotics Screening in Urine
An Automated Procedure for Determining ITSP SPE Analytical Parameters for the Analysis of Cortisol and Cortisone in Urine


ITSP Hardware Kits and Accessories

ITSP Hardware Kits offer you the proper hardware components to manage ITSP. Included in each kit is the Prep Tray, Waste Receptacle, a Needle Guide and a replacement screw for the LC inlet.
If you are using ITSP for Solid Phase Extraction you may also need additional software, solvent reservoirs, a fast wash station or other sample trays. In this section you will see a variety of solutions for these additional needs.

 ↳ leaptec.com  ↳ leapwiki.com

LEAP provides automated workstation instrumentation solutions based on the LEAP CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography.LEAP offers full support and service for the PAL platform in addition to being able to write custom macros, cycles, and scheduling to your applications. Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions.

Contact LEAP

Contact the LEAP office

Personal tools