Hydrogen Deuterium Exchange

From LEAP

Revision as of 15:24, 22 June 2009 by Bpeat (Talk | contribs)
Jump to: navigation, search
Hydrogen Deuterium Exchange
Application Type
  SPECIAL
Application ID
  HD-X PAL
Description
  Automated Sample Prep for HD-X

Contents

General Description

Hydrogen Deuterium Exchange (HD-x) is an experimental technique to obtain structural data on proteins. It relies on the accurate measurement of the degree of labeling of the protein by deuterated hydrogen during a precisely measured labeling interval. The labeling reaction is stopped by addition of a quenching reagent, and loss of label between the quench step and analysis (usually by LCMS) is minimized by keeping the reactants at a temperature close to zero degrees Celsius.

The technique is difficult to perform manually due to the demands for precise timing and reproducing exactly the same experimental conditions. Automation solves both of these challenges. LEAP has developed a robotic solution to the sample Prep and introduction to the LCMS. It consists of a Twin PAL workstation configured as shown below. There are two cooled zones on the workstation: A cooled stack at 1oC for the samples and labeled reactants and a cooled chromatography module which contains the injection valve, the column selection valves and the columns themselves. These are maintained also at around 1degree C. The automation process is controlled using LEAP Shell software.

Significant Markets

  • Pharmaceutical - Drug Discovery
  • Academia - Protein structural studies

Protein Structure

Protein function is strongly related to their tertiary structure and conformation. Chemical interactions which change this conformation can significantly affect protein functionality. This is important in studying disease states or effectiveness of drug candidates.

There are 3 commonly used techniques to study protein structure:

1. Crystallography

2. NMR

3. HD Exchange

HD Exchange is a good approach in cases where there is insufficient sample (NMR) or the protein will not crystallize. It relies on the accurate measurement of the degree of labeling of a protein by deuterated hydrogen during a precisely measured labeling interval. The deuterium is exchanged with hydrogen atoms on the exposed amide backbone of the protein and depending on the which regions are exposed and how the protein is folded, differences are seen in the MS spectra of the digested protein when compared with non-labeled proteins.

Labeling time is important (exposure time of the protein to Deuterium) as is prevention of "back-exchange" after the labeling has occurred. This is the natural tendency of normal hydrogen to replace the deuterium after the label is removed. This is minimized by reducing the temperature of the mixture to close to zero degrees Celsius.

What is an HD-X experiment ?

Performing HD-X experiments manually involves setting up many labeling mixtures then quenching these mixtures at precisely timed intervals (by changing the pH) and then injecting them onto a chilled injection valve to the LCMS. In most cases a range of label times is used, from 10 seconds to several hours. Sources of error (apart from normal liquid handling inaccuracies) are timing errors, and degradation of sample due to "back-exchange". All of these sources of error are eliminated using the automated system described here.

Overview of H/D-x PAL System

H/D-X PAL™ is an easy-to-use, system that provides an automated process for the scheduling and experimental execution of H/D-X experimental workflow. By use of the advanced LEAP Shell scheduling software experimental design is simplified and reliable. Synchronous reagent addition and sample labeling steps are automatically scheduled to increase throughput and produce high quality data.

What does the H/D-X PAL do?

Image:LineDrawing HD-x system.png

The instrument has two cooled zones for sample storage and labeling. The injection valve is chilled to 1 degree Celsius and includes a solvent pre-chiller and column selection valve for in-line protein digestion.

A typical experiment would consist of around a dozen time points for each protein sample, ranging from 10 seconds to 8 hours. The PAL system will schedule the experiments and perform each of the following 3 steps for each time point and sample:

Labeling The protein is mixed with a Deuterium reagent and incubated for a precisely timed interval at room temperature.

Quenching The labeling reaction is stopped by transferring a volume of the mixture to an excess of quenching reagent in a pre-cooled vessel.

Injection A measured portion of the quenched reactants is injected into a multi-column, multi-valve LCMS system. This step is synchronized with the analytical system using hardware signals. The PAL will inject samples as the LCMS system becomes ready. Labeled and quenched reactants are held in the cooled stack until they have been injected and analyzed.


Software Control

LEAP has developed a scheduling application using their proprietary software “LEAP Shell” which will automate the above labeling steps, as well as the injection to the MS and the associated chromatography.


Image:Scr.Shot_-_Sample_List.png

LEAP Shell Screen Shot 1: User interface to set up the labelling experiment:


Image:Scr.Shot_-_PAL_Method.png

LEAP Shell Screen Shot 2: Powerfull editor for the PAL Methods:


Image:Scr.Shot_-_QueueList.png

LEAP Shell Screen Shot 3: Run-time display of the labelling experiment with actual times:



Useful links

HXMS.com A good general reference site on HD Exchange theory and practice

ASMS Interest Group - Hydrogen Exchange and Covalent Labeling

LEAP Website H/D-x product description

Publications

American Chemical Society 2006. Chalmers
Poster from ASMS 2006
Poster ASMS 2006 Washington Univ. In St Louis

Other Reference material

LEAP HDX Application Note
Collaboration between LEAP & ExSAR
Comparison of Data Processing software


This workstation application is based on the CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your LC or GC requirements. Please contact LEAP Technologies on how we can help you get maximized throughput with flexible automation.

Contact LEAP

Contact the LEAP office




Personal tools