M-001 Dual Channel Serial Injection - Analyst with ICC-CE
From LEAP
(Difference between revisions)
Line 1: | Line 1: | ||
{{LSApp | {{LSApp | ||
|name = LC Column Switching & Regeneration | |name = LC Column Switching & Regeneration | ||
- | |image = | + | |image = 2_valve_on_top.png |
|type = '''SPECIAL''' | |type = '''SPECIAL''' | ||
|id = M-001 | |id = M-001 | ||
Line 8: | Line 8: | ||
=== Overview === | === Overview === | ||
- | This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This | + | This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within '''Analyst''' using a Custom LEAP Cycle (requires ICC-CE software). |
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.<br> Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle. | There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.<br> Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle. |
Revision as of 17:06, 25 March 2009
LC Column Switching & Regeneration | |
Application Type | |
SPECIAL | |
Application ID | |
M-001 | |
Description | |
LC Column Switching & Regeneration. |
Overview
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within Analyst using a Custom LEAP Cycle (requires ICC-CE software).
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.
Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.
Contact LEAP
Keyword: Analyst Software, ICC-CE, PAL, 10 port valve, selector valve, injector valve, gradient pump, regeneration pump,back cut, peaks of interest, GLP