LEAP X-Press Filtration Station

From LEAP

(Difference between revisions)
Jump to: navigation, search
(Accessories for the PAL)
(Accessories for the PAL)
Line 90: Line 90:
[[image:Accessories_Icon.png|40px]] [http://www.leapwiki.com/mediawiki/index.php?title=Category:Accessories Other Accessories for PAL Robots]
[[image:Accessories_Icon.png|40px]] [http://www.leapwiki.com/mediawiki/index.php?title=Category:Accessories Other Accessories for PAL Robots]
<br><br>
<br><br>
-
'''Materials of Construction of Whatman UniPrep Vials '''
+
'''Materials of Construction of Whatman UniPrep Vials '''<br>
-
Housing and cap --> Polypropylene
+
Housing and cap --> Polypropylene<br>
-
Septa --> PTFE coated silicone rubber
+
Septa --> PTFE coated silicone rubber<br>
<br>
<br>
'''Starting point for what filtering media material may work best:'''
'''Starting point for what filtering media material may work best:'''

Revision as of 14:27, 25 February 2010

Sample Filtration
Application Type
  SPECIAL
Application ID
  LEAP X-Press
Description
  Sample Filtration with LEAP X-Press Individual Filter System
 ↳ leaptec.com  ↳ leapwiki.com

Contents

Overview

Whatman Mini-UniPrep™ Syringeless Filters offer a convenient way of filtering samples containing particulates prior to introduction on an HPLC system. It combines filtration and storage into one simple device that can be integrated with the PAL. They currently come with 2 pore sizes and many different filter media. They are designed for manual operation but are not conducive to large batch sizes - repeated pressing by hand being very fatiguing.
Mini UniPrep Syringeless Filters are specifically designed to remove particulates from smaller sample volumes. Mini-UniPrep consists of a 0.4 ml capacity chamber and a plunger. The plunger contains a filtration membrane at one end and a pre-attached cap/septum at the other. The plunger is pressed through the sample in the outer chamber and positive pressure forces the filtrate into the reservoir of the plunger. Air escapes through the vent hole until the locking ring is engaged providing an air-tight seal. (see Diagram)

LEAP has developed a "Press" accessory,LEAP X-Press, for the PAL which allows samples to be filtered reliably and repeatably. It does not require external pneumatics, but compress the filter chambers. Samples can be processes in batch mode or "just in time" before direct injection to the LC valve. The LEAP X-Press can be used for stand-alone operation with an optional foot.


Image:UniPrep_Filter_diagram.png

The PAL offers precise control of the down force on the sample. The user can specify the speed of compression in the method, allowing even very heavy suspensions to be filtered successfully.

The complete system can provide automated online filtration, including sample preparation, sample transport and injection into the chromatographic system.


Info on LEAP X-Press

Features and Benefits

  • Ease of use Filtering and storage vial allows you to process sample loads in one-third the time
  • Wide range of membranes Suits a variety of sample types
  • Available in 0.2 and 0.45 μm pore sizes to meet specific sample application requirements
  • Compatible with all major autosamplers Allows you to speed processes
  • Fewer consumables required Lowers Costs by up to 40 percent
  • Slit septa cap - Enables Mini-UniPrep use with current robotics on HPLC instruments for high throughput automation
  • Slit septa cap - Increases needle longevity Eliminates coring problems associated with repeated sampling
  • Slit septa cap - Durable yet flexible and has been specially designed for instruments with sensitive sampling needs

The self-sealing septa of the UniPrep allows for repeated injections: Prevents evaporation and maintains sample integrity. Whatman Mini-UniPrep Syringeless Filters offer a highly efficient, simple alternative for removing unwanted protein prior to HPLC/MS analysis. Ideal for performing analytical characterization in drug research, the method utilizes acetonitrile precipitation and filtration by compression to remove protein from plasma, serum, whole blood, and other biological fluids. It is a single tube method that saves time and eliminates the manual transfer steps which make spin clarification problematic.

Significant Markets

Any labs who are doing High Performance Liquid Chromatography (HPLC) or more significantly, Ultra High Performance Liquid Chromatography (UHPLC) analysis. It is imperative that particles are eliminated to avoid blockage of HPLC lines and columns. This filtering technique let you prepare samples in one third the time required by other methods. Automation of the filtering step has NOT been possible in a reliable fashion till now.

Filtration on a single disposable for a great variety of situations.

  • Samples can be filtered just before injection onto a UHPLC system.
  • Plant extracts and any solutions with particles

Possible Markets

  • Cosmetics
  • Food & Beverage
  • Pharmaceutical
  • Routine analysis
  • Composite assays
  • Content uniformity
  • Protein precipitation
  • Solubility testing
  • Dissolution testing
  • Sample filtration

Image:XPress_on_PAL.png

Filtration is in general, the removal of parts from a “raw” sample when passing through a filtration material.

  • Prevents blocking of capillaries, frits, and the column inlet
  • Results in less wear and tear on the critical moving parts of injection valves
  • Produces improved analytical results by removing potentially interfering contamination
  • Results in less downtime of the instrument for repairs


Photos

Print this page

Print this page

More Info

Whatman Uniprep Website
Whatman UniprepSellSheet

Videos

Movie of the Press in action on YouTube

Various PAL's in action on YouTube

Accessories for the PAL

Other Accessories for PAL Robots

Materials of Construction of Whatman UniPrep Vials
Housing and cap --> Polypropylene
Septa --> PTFE coated silicone rubber

Starting point for what filtering media material may work best:

  • General filtration media/Solvent based samples --> Polypropylene (PP)
  • Particulate laden liquids --> Glass Microfiber (GMF)
  • Aqueous/Organic samples in 3 to 10 pH range --> Nylon (NYL)
  • Chemically aggressive solutions --> Polytetrafluroethylene (PTFE)
  • Biological samples requiring low protein binding media --> Polyethersulfone (PES) or Regenerated Cellulose (RC)
  • Aqueous/Organic solvents – low non-specific protein binding media --> Polyvinylidenefluoride (PVDF)
  • Aqueous/Organic solvents – high flow and loading capacity --> Polypropylene Depth (dp PP)



 ↳ leaptec.com  ↳ leapwiki.com

LEAP provides automated workstation instrumentation solutions based on the LEAP CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography. LEAP offers full support and service for the PAL platform in addition to being able to write custom macros, cycles, and scheduling to your applications.Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions.

Contact LEAP

Contact LEAP

For additional information about LEAP and the PAL Platform, please contact LEAP Technologies.

Keywords

Whatman Mini-Uniprep syringeless filters,eliminate syringe filter, Amber Housing (for Light Sensitive Samples),simple to use protection, single step filter vials, plunger filter, simultaneously filtering and readying sample, fast, compatible, debris blockage HPLC, All-in-one filtration process, pre-slit septa, Nylon, PES, PVDF ,RC,PTFE, PP,GMF, Nylon, PES , PVDF, Thomson , Equivalent in size to 12 x 32 mm vials,

Personal tools