Micro Extraction by Packed Sorbent

From LEAP

(Difference between revisions)
Jump to: navigation, search
Current revision (00:45, 21 October 2010) (view source)
 
(27 intermediate revisions not shown.)
Line 6: Line 6:
   |description = Micro Extraction by Packed Sorbent (MEPS)  
   |description = Micro Extraction by Packed Sorbent (MEPS)  
}}
}}
-
[[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]]
+
{{logo}}
-
 
+
=== Overview ===
=== Overview ===
Line 26: Line 25:
MEPS for PAL Systems is distributed by SGE & LEAP Technologies.
MEPS for PAL Systems is distributed by SGE & LEAP Technologies.
----
----
-
'''Benefits of MEPS''
+
'''Benefits of MEPS'''
* MEPS allows SPE methodology to be applied to small sample volumes.
* MEPS allows SPE methodology to be applied to small sample volumes.
* MEPS can be integrated into autosampler robotics and allows on-line use of SPE.
* MEPS can be integrated into autosampler robotics and allows on-line use of SPE.
Line 42: Line 41:
Like conventional SPE, the number of times the device can be reused is
Like conventional SPE, the number of times the device can be reused is
dependent on the sample matrix. For simple applications, MEPS devices have
dependent on the sample matrix. For simple applications, MEPS devices have
-
been used successfully for >50 cycles.
+
been used successfully for >50 cycles.<br><br><br>
[[Image:Process_of_MEPS.png]]
[[Image:Process_of_MEPS.png]]
----
----
Line 75: Line 74:
   Image:Process_of_MEPS.png | Process of MEPS
   Image:Process_of_MEPS.png | Process of MEPS
   Image:MEPS Kit.png | MEPS Kit
   Image:MEPS Kit.png | MEPS Kit
-
  Image:MEPS-Web.pdf | Information specific to MEPS for CTC Autosamplers MEPS Online – The SPE solution for your CTC platform
 
</gallery>
</gallery>
 +
 +
=== Publications ===
 +
[[image:Download Icon.png|40px]] [[Media:MEPS-Web.pdf | Information specific to MEPS for CTC Autosamplers MEPS Online – The SPE solution for your CTC platform]]<br>
 +
 +
----
 +
=== MEPS Applications ===
 +
[[image:Download Icon.png|40px]] [http://www.sge.com/uploads/dd/ec/ddecaa2dc33f176f23d341464c0b695b/TP-0180-M1.pdf APPLICATION OF MEPS COUPLED WITH GAS CHROMATOGRAPHY TIME-OF-FLIGHT MASS SPECTROMETRY (GC-TOF MS) IN ANALYSIS OF BROMINATED FLAME RETARDANTS IN WASTE WATER]
 +
 +
[[image:Download Icon.png|40px]] [http://www.sge.com/uploads/66/7d/667ded5291cfd6aa36baf3960f4b7cf1/TP-0142-C.pdf Application to Opium Contaminated Animal Feeds]
 +
 +
[[image:Download Icon.png|40px]] [http://www.sge.com/uploads/b9/5a/b95a6c7492cba44b236e6eda3c4718cb/TP-0161-M.pdf ON-SITE SAMPLE PREPARATION USING MEPS FOR WASTE WATER ANALYSIS]
 +
 +
[[image:Download Icon.png|40px]] [http://www.sge.com/uploads/af/ae/afaed573d1dd9cb7f72e12dce019650a/AN-0032-M.pdf FALSE POSITIVE DRUG TESTS FROM CONTAMINATED FEED]
 +
 +
[[image:Download Icon.png|40px]] [http://www.sge.com/uploads/dd/ec/ddecaa2dc33f176f23d341464c0b695b/TP-0180-M1.pdf ANALYSIS OF BROMINATED FLAME RETARDANTS IN WASTE WATER]
----
----
Line 82: Line 95:
=== More Info ===
=== More Info ===
-
[http://www.sge.com/products/meps MEPS web site by SGE including product brochure]
+
[[image:Download Icon.png|40px]] [http://www.sge.com/products/meps MEPS web site by SGE including product brochure]
 +
[[image: Info icon.png|40px]] [http://www.nxtbook.com/nxtbooks/advanstaruk/thecolumn0707/index.php?startid=26 The Column article "It's a Small World"]
 +
[[image: Info icon.png|40px]] [http://www.nxtbook.com/nxtbooks/advanstaruk/thecolumn0408/index.php?startid=12 The Column article "On-Site Sample Preparation Using MEPS for Wastewater Analysis ]
 +
 +
=== MEPS Online Kit Application Index ===
 +
<br>
 +
''ALKALOIDS''
 +
* A01 3,4-Diaminopyridine from Urine (Lambert-Eaton)
 +
* A02 Alkylbenzyl Dimethylammonium Chlorides from Plasma
 +
* A03 Amino Acids from Urine
 +
* A04 Basic Nucleosides from Aqueous Solution
 +
* A05 Bio-fl avonoids from Red Wine
 +
* A06 Casuarine in Plant Extract
 +
* A07 Catecholamines from Urine
 +
* A08 Cyclodextrins from Plasma
 +
* A09 Cyclodextrins from Urine
 +
* A10 Dipterex in Serum
 +
* A11 Diterpene Glycosides from Tea Extract
 +
* A12 Fatty Acids from Serum
 +
* A13 Homovanillic Acid from Plasma
 +
* A14 Isofl avones from Plant Extract
 +
* A15 Mixed Aromatic Amines from Urine and Plasma
 +
* A16 Nicotine and Cotinine from Plasma
 +
* A17 Nucleosides from Aqueous Solution
 +
* A18 Nucleosides from Plasma
 +
* A19 Nucleosides from Urine
 +
* A20 PEG 400 in Serum
 +
* A21 Persistent Organic Pollutants PAH, PCB and Pesticides in Blood
 +
* A22 Persistent Organic Pollutants PAH, PCB and Pesticides in Plasma
 +
* A23 Persistent Organic Pollutants PAH, PCB and Pesticides in Serum
 +
* A24 Pesticides and PCB in Fats
 +
* A25 Phenylanaline from Plasma
 +
* A26 Prostaglandins from Urine
 +
* A27 Prostaglandins from Whole Blood
 +
* A28 Salsoline from Plasma
 +
* A29 Steroid Acids from Serum
 +
* A30 Tryptophan from Plasma
 +
* A31 Vanillymandelic Acid from Plasma
 +
* A32 Xanthines (Caffeine) from Serum
 +
* A33 Xanthines (Theophylline) from Serum
 +
<br>
 +
''ENVIRONMENTAL''
 +
* E01 Carbamate Insecticide (Aldicarb) from Water
 +
* E02 PAH and PCB in Contaminated Soi
 +
* E03 Phenols in water
 +
* E04 Phthalate Esters in Water
 +
* E05 S-triazine Herbicide (Atrazine) in Soil
 +
<br>
 +
''FOOD & BEVERAGE''
 +
* F01 Afl atoxin B2 and M2 Metabolite Trace Analysis in Milk
 +
* F02 Chloroacetanilide Herbicides (Acetochlor and Metolachlor )in Contaminated Water
 +
* F03 F-2 Mycrotoxin Trace Analysis in Cereal
 +
* F04 Fatty Acid Methyl Esters (Long Chain) in Fermentation medium
 +
* F05 Omega 6 Fatty Acid in Malt Lipid Fractions
 +
* F06 Pigment Anthocyanidins in Wine
 +
* F07 S-triazine Herbicide (Atrazine) in Cereal
 +
* F08 Sulfonamide Trace Analysis in Meat
 +
<br>
 +
''PHARMACEUTICALS''
 +
* P01 Acetazolamide and Bumetanide in Urine
 +
* P02 Acetazolamide in plasma
 +
* P03 Amiodarone, Fendiline and Procainamide in Serum
 +
* P04 Amphetamine in Plasma
 +
* P05 Analgesics in Serum; Paracetamol and Tramadol
 +
* P06 Anesthetics in Serum; Benzocaine, Mepivacaine, Procaineand Lidocaine
 +
* P07 Antidepressants (Trycyclics) in Blood
 +
* P08 Antidepressants (Trycyclics) in Urine
 +
* P09 Atenolol in Plasma
 +
* P10 Barbiturates in Serum; Barbital, Amobarbital, Phenobarbital,Secobarbital
 +
* P11 Barbiturates in Urine; Barbital, Amobarbital, Phenobarbital,Secobarbital
 +
* P12 Carbamazepine, Phenobarbital and Primidone in Serum
 +
* P13 Chloramphenicol in Eye Drops
 +
* P14 Cimetidine in Plasma
 +
* P15 Cyclosporin in Blood
 +
* P16 Erythromycin and Clarithromycin in Urine
 +
* P17 Minor Tranquilizers (Benzodiazepines) in Urine or Serum
 +
* P18 Minor Tranquilizers (Diazepam and Lorazepam) in Hair
 +
* P19 Opiate Analgesics in Blood; Morphine and Codeine
 +
* P20 Propranolol in Serum
 +
* P21 Stobadin from Serum
 +
* P22 Vitamin A, D and E in Supplements
 +
* P23 Vitamin D3 and Metabolites in Serum
----
----
-
[[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]]
 
-
=== Contact LEAP ===
 
-
For additional information about this technique please contact LEAP Technologies for detailed information
+
On-Line and Off-Line Application Of Micro-SPE (MEPS)
-
* [http://www.leaptec.com/contact-us.php Contact the LEAP office]
+
by Peter Dawes, Ern Dawes, Paul Wynne, SGE Analytical Science, and Dan DiFeo, SGE Inc.
 +
Introduction
 +
Solid-phase extraction (SPE) has revolutionized sample preparation. Variations on the technique offer enhanced recovery, greater speciation and reduced solvent and sample consumption over other techniques. Micro-Extraction Packed Sorbent (MEPS) is the miniaturization of conventional SPE from milliliter to microliter bed volumes that allows SPE to be used with very small samples. The manipulation of the small volumes is achieved with a precision gas tight syringe. With a typical void volume of 7 L, the MEPS elution is compatible with GC and LC inlets making it ideal for integration into an automated sampling system for online SPE.
 +
In most cases, MEPS allows the same level of sample concentration as is possible with off-line conventional SPE while providing opportunities for truly hybrid multi-dimensional methods. MEPS methods may be readily adapted from established SPE methods including those based on mixed mode or complex chemistries.
 +
 +
Figure 1. MEPS needle with integral 45-m C18 packed bed SPE device (void volume approximately 7 L). Click to enlarge
 +
Like SPE, MEPS is for use with liquid samples (either normal or reversed phase) and yields four fractions: the unretained, weakly bound, strongly bound and irreversibly bound. However, because MEPS is a double pass system (sample and solvent enter and exit from the bottom of the bed), the weakly bound fraction (commonly the interferences eliminated by washing) is less strongly bound. The irreversibly bound fraction affects MEPS and conventional SPE and is usually associated with sorbent wetting rather than sample purification and so the irreversible binding of matrix material from one sample does not preclude reuse of the device for a sample of the same type.
 +
 +
Like conventional SPE, the number of times the device can be reused is dependent on the sample matrix. For simple applications, MEPS devices
 +
 +
have been used successfully for 50 cycles.
 +
 +
Benefits of MEPS:
 +
 +
MEPS allows SPE methodology to be applied to small sample volumes.
 +
MEPS can be integrated into autosampler robotics and allows online use of SPE.
 +
MEPS can reduce sample and reagent consumption and waste disposal.
 +
Double pass flows can reduce the weakly bound fraction.
 +
MEPS is field portable for remote sampling with or without the use of automated equipment.
 +
MEPS is adaptable for other analytical techniques including immunoassay and off-line analysis by NMR, IR and other methods.
 +
 +
Plant Extracts
 +
Figure 2. Phytolacca octandra.
 +
The plant Phytolacca octandra (Figure 2) is one of the inkweeds or Pokes. In this application, the aerial portions of the plant were homogenized in acidified methanol, allowed to percolate for 12 hours, filtered and then extracted using a C2 MEPS cartridge. The MEPS was conditioned with methanol (30 L), water (30 L) and then 100 L of the plant extract was passed through the sorbent at 5mL/sec. The exhausted fraction was ejected at the same rate and the sorbent washed with 100mL of water. The sorbent was dried with air (3 80 L at 50 mL/sec) and eluted sequentially with hexane (10 L), dichloromethane (10 L) and methanol (10 L). The eluates were analyzed directly by GCMS on a BPX5 column (Figure 3).
 +
 +
The C2 MEPS method allowed the one step isolation of the FAME fraction (hexane) and the elimination of the highly polar sugar fraction (see Figure 3). Speciation of polar and non-polar analytes from a single sample digest was readily achieved without the need for off-line sample preparation.
 +
 +
Waste Water
 +
To demonstrate the usefulness of MEPS for dilute samples with a relatively simple matrix, a surrogate waste water sample was prepared from clear phenol free waste water spiked with either 25-ppb or 250-ppt phenols. The water was extracted with a C18 MEPS cartridge conditioned with methanol (30 L), water (30 L) and then 10 100 L of the water at 5 L/sec. The exhausted water was ejected at the same rate after each cycle; the sorbent was air dried (3 80 L at 50 L/sec). The analytes were eluted with methanol (10 L) and the fraction analyzed directly by GCMS on a BPX5 column. Chromatograms for the water samples are shown in Figures 4 and 5. Carryover was examined following the extraction of the 25-ppb sample by elution of a second and third portion of methanol without any intervening wash steps (Figure 5).
 +
 +
Figure 3. Effectiveness of the water wash in a GCMS analysis of Phytolacca octandra extract before and after fractionation on MEPS C2. Click to enlarge.
 +
Figure 4. Phenols in waste water at 250 ppt. Click to enlarge.
 +
Figure 5. Phenols in waste water at 25 ppb. Click to enlarge.
 +
 +
The C18 MEPS method allowed the one step isolation of phenols from water with good recovery, linearity and little carryover. The sorbent was reusable for the application for a large number of samples with no loss of performance after 10 analyses.
 +
 +
Toxicology Samples
 +
Papaver somniferum (opium poppy) is a feed contaminant that can result in positive drug tests for racing horses. We describe the extraction of a urine sample from an animal receiving contaminated feed as a demonstration of the off-line application of mixed mode C8/SCX MEPS for complex biological fluids.
 +
 +
Figure 6. Opiate metabolites in urine following MEPS of hydrolyzed urine. Click to enlarge.
 +
Here, 300 L of a diluted equine urine was hydrolyzed with b-glucuronidase or acid, filtered and extracted on a C8/SCX MEPS cartridge conditioned with methanol (30 L), potassium phosphate buffer (0.2 M, pH 6, 30 L) at a flow rate of 5 L/sec. The exhausted fraction was ejected at the same rate and the sorbent washed with 100-L phosphate buffer, 50-L acetic acid (1% v/v) and 100-L methanol. The sorbent was dried with air (3 80 L at 50 L/sec) and the sorbent eluted with 20-L dichloromethane-isopropanol-ammonia (49:49:2). The organic phase was evaporated under nitrogen and [http://www.leapwiki.com/mediawiki/index.php?title=Category:Derivatization derivatized] with 10 L of acetic anhydride-pyridine (1:2) at 80 C for 30 minutes before evaporation and reconstituted in 5 L of ethyl acetate. The extract was analyzed by GCMS on a BPX5 column (Figure 6).
 +
 +
The C8/SCX MEPS method allowed the microscale preparation of a small volume sample with comparable performance to conventional SPE techniques. Used off-line with [http://www.leapwiki.com/mediawiki/index.php?title=Category:Derivatization derivatization] and GCMS here, the sample was also suitable for online ESI-LCMSMS analysis by changing the elution solvent to methanol-ammonia (98:2) or methanol-trimethylamine (98:2).
 +
 +
For more information, contact Dan DiFeo, technical support, SGE Inc., at ddifeo@sge.com or by phone at 800-945-6154. Company’s Other Products
 +
Laboratory Equipment
 +
Rockaway NJ 07866
 +
 +
=== Videos of PAL ===
 +
<br>
 +
[[Image:Movie Icon.png|40px]]
 +
[http://www.youtube.com/user/LEAPTechnologies#g/u LEAP's PAL Application Videos on YouTube]
 +
<br>
 +
----
 +
[[Image:MEPS.png|center|1000px]]<br><br>
 +
 +
----
 +
{{logo}}
 +
 +
[http://www.leaptec.com/ LEAP Technologies] provides fully automated sample processing workstation instrumentation solutions based on the LEAP CTC [http://leapwiki.com/mediawiki/index.php?title=Category:PAL PAL] X, Y, Z syringe only autosampler robot. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography. LEAP also offers process control systems and stand alone workstations. LEAP is an open, neutral resource for technical support of existing PAL. Keep your PAL in peak working order LEAP offers full [http://www.leapwiki.com/mediawiki/index.php?title=Category:Service support and service] for the PAL platform with over 11 years of history with the PAL. LEAP can also write custom macros, [http://www.leapwiki.com/mediawiki/index.php?title=Category:Cycles cycles], and scheduling [http://www.leapwiki.com/mediawiki/index.php?title=Category:Software software] for your specific applications. We're proud that LEAP is known as a productivity improver for laboratories including government labs in agencies such as FDA, USDA, EPA, FBI, Armed Forces, Homeland Security and others function more efficiently. Robotic sample prep equipment works well in government labs and challenges lab technicians to keep feeding those more samples. Analytical equipment such as chromatographs and mass spectrometers are expensive devices and also expensive to maintain. LEAP can help to keep them running and analyze samples around the clock. LEAP has most of its laboratory automation robots under [http://leapwiki.com/mediawiki/index.php?title=Category:GSA GSA]contract and is eager to service and maintain everything it sells to keep your expensive lab equipment running around the clock. CTC Analytics Combi PAL and HTS PAL for Front-End Automation for LC and GC applications. LEAP offers customized on-site training. LEAP Is continuing to develop FAQ, Solutions, Tips & Techniques,a Learning Center with the LEAP Wiki for the benefit of all PAL Platform Robotic users. We will continue exceeding customer expectations for quality, cost effective and efficient results using the most current technology and expertise in our industry. Our aim is to be the ultimate knowledge base on everything related to the PAL. LEAP is known as the autosamplerguys being in business over twenty years. LEAP has extensive multi-vendor experience with integrating the PAL. HTX, HTS, HTC, HTX-xt, HTS-xt, HTC-xt, LC-Mini, Combi PAL,COMBI-xt PAL, and GC PAL models. We have the Quality, Training, Parts availability, Infrastructure, and Experience.LEAP is sometimes referred to as an Application Automation House who is continually adding new products and solutions to its Lab Automation line. As we do so we change the content of this web site so please check back.<br>Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions whether it is our standard applications or customized to your needs. The Combi-xt PAL autosampler is able to perform liquid and headspace analysis. Headspace options include Solid Phase Micro-extraction (SPME) and ITEX-2 as alternative to purge and trap. The HTS-xt PAL autosampler dramatically increases sample throughput by offering automated sample prep, Peltier cooled and heated sample storage using the cooled stack, Peltier cooled and heated sample storage using the cooled stack and low carryover due to the self-cleaning DLW (Dynamic Load and Wash) module and L-Mark syringes, peltier cooled valves and columns.  LEAP provides automation that fits your needs and reduces errors, increases sample capacity, increases efficiency, reduces cost by supplying long term productivity improvements with user friendly software and flexible hardware. LEAP performs just in time or on time sample prep and inject. LEAP offers full support of integrated and non-integrated PAL software with both sales and support for Cruise Control, Cycle Composer and Cycle Composer Macro Editor, Agilent ChemStation CTC Control, Mass Hunter and EZChrom CTC integration, Xcalibur, Waters Masslynx or Analyst CTC Control software, Thermo (Finnegan) XCalibur, LEAP Shell, and Chronos software. The PAL is more than just an autosampler, it is a high-quality and efficient laboratory robot platform that can be customized by LEAP Technologies to meet your specific needs.
 +
 +
=== Contact LEAP ===
 +
{{contact|topic=LEAP and the PAL Platform}}
 +
<br>''The information and references contained in this web site are believed to be accurate and the most current information available.''
-
<br><br>
 
[[Category:Accessories]]
[[Category:Accessories]]
[[Category:SPE]]
[[Category:SPE]]
 +
[[Category:Application_Solutions]]

Current revision

SPE Applications
Application Type
  STANDARD AND SPECIAL
Application ID
  MEPS
Description
  Micro Extraction by Packed Sorbent (MEPS)
 ↳ leaptec.com  ↳ leapwiki.com

Contents

Overview

Micro Extraction by Packed Sorbent* (MEPS) is a new development in the field of sample preparation by solid phase extraction (SPE), and is a result of a collaboration between SGE and Mohamed Abdel-Rehim (AstraZeneca) and Lars G. Blomberg (University of Karlstead) Automation of MEPS is accomplished with the PAL Series of CTC Autosamplers. MEPS is the miniaturization of conventional SPE packed bed devices from milliliter bed volumes to microliter volumes. The MEPS approach to sample preparation is suitable for reversed phases, normal phases, mixed mode or ion exchange chemistries. MEPS is available in a variety of common SPE phases. The cartridge (Patent Pending) contains the stationary phase, and is built into the syringe needle. With a typical void volume of a few μL, the MEPS elution is compatible with GC and LC inlets making it ideal for automated on-line SPE on the CTC PAL platform. MEPS kits are exclusively available worldwide through LEAP Technologies and SGE Analytical Science.


MEPS performs the same function as SPE, namely the purification or speciation of samples, but with some significant differences:

  • MEPS works with much smaller samples (as small as 10µL) than full scale SPE
  • MEPS can be fully automated – the sample processing, extraction and injection steps are performed on-line using the same syringe
  • MEPS is applicable to GC and LC
  • Significantly reduces the volume of solvents and sample needed

MEPS online is an SPE solution for rapid development, in a variety of matrices; plasma, serum, whole blood, urine, aqueous


MEPS for PAL Systems is distributed by SGE & LEAP Technologies.


Benefits of MEPS

  • MEPS allows SPE methodology to be applied to small sample volumes.
  • MEPS can be integrated into autosampler robotics and allows on-line use of SPE.
  • MEPS can reduce sample and reagent consumption and waste disposal.
  • Double pass flows can reduce the weakly bound fraction.
  • MEPS is field portable for remote sampling with or without the use of automated equipment.
  • MEPS is adaptable for other analytical techniques including immunoassay and off-line analysis by NMR, IR and other methods.

Solid-phase extraction (SPE) has revolutionized sample preparation. Variations on the technique offer enhanced recovery, greater speciation and reduced solvent and sample consumption over other techniques. Micro-Extraction Packed Sorbent (MEPS) is the miniaturization of conventional SPE from milliliter to microliter bed volumes that allows SPE to be used with very small samples. The manipulation of the small volumes is achieved with a precision gas tight syringe. With a typical void volume of 7μL, the MEPS elution is compatible with GC and LC inlets making it ideal for integration into an automated sampling system for on-line SPE. In most cases, MEPS allows the same level of sample concentration as is possible with off-line conventional SPE while providing opportunities for truly hybrid multi-dimensional methods. MEPS methods may be readily adapted from established SPE methods including those based on mixed mode or complex chemistries.
Like SPE, MEPS is for use with liquid samples (either normal or reversed phase) and yields four fractions: the unretained, weakly bound, strongly bound and irreversibly bound. However, because MEPS is a double pass system (sample and solvent enter and exit from the bottom of the bed, the weakly bound fraction (commonly the interferences eliminated by washing) is less strongly bound. The irreversibly bound fraction affects MEPS and conventional SPE and is usually associated with sorbent wetting rather than sample purification and so the irreversible binding of matrix material from one sample does not preclude reuse of the device for a sample of the same type.
Like conventional SPE, the number of times the device can be reused is dependent on the sample matrix. For simple applications, MEPS devices have been used successfully for >50 cycles.


Image:Process_of_MEPS.png


Available for both GC & LC applications. There are over 65 pre-set methods.

Significant Markets

  • Natural Products
  • Environmental
  • Food & Beverage
  • Pharmaceuticals

Sample Methods

  • Aflatoxin B2 and M2 Metabolite
  • Amphetamine
  • Catecholamines
  • Cyclosporin
  • F-2 Mycotoxin
  • Opiate Anagelesics
  • Phthalate Esters
  • Prostaglandins
  • Persistent Organic Pollutants (PAH, PCB, and Pesticides)
  • Steroid Acids
  • S-triazine Herbicide (Atrazine)
  • Vitamin A, D, and E


Photos

Publications

Information specific to MEPS for CTC Autosamplers MEPS Online – The SPE solution for your CTC platform


MEPS Applications

APPLICATION OF MEPS COUPLED WITH GAS CHROMATOGRAPHY TIME-OF-FLIGHT MASS SPECTROMETRY (GC-TOF MS) IN ANALYSIS OF BROMINATED FLAME RETARDANTS IN WASTE WATER

Application to Opium Contaminated Animal Feeds

ON-SITE SAMPLE PREPARATION USING MEPS FOR WASTE WATER ANALYSIS

FALSE POSITIVE DRUG TESTS FROM CONTAMINATED FEED

ANALYSIS OF BROMINATED FLAME RETARDANTS IN WASTE WATER


More Info

MEPS web site by SGE including product brochure

The Column article "It's a Small World"

The Column article "On-Site Sample Preparation Using MEPS for Wastewater Analysis

MEPS Online Kit Application Index


ALKALOIDS

  • A01 3,4-Diaminopyridine from Urine (Lambert-Eaton)
  • A02 Alkylbenzyl Dimethylammonium Chlorides from Plasma
  • A03 Amino Acids from Urine
  • A04 Basic Nucleosides from Aqueous Solution
  • A05 Bio-fl avonoids from Red Wine
  • A06 Casuarine in Plant Extract
  • A07 Catecholamines from Urine
  • A08 Cyclodextrins from Plasma
  • A09 Cyclodextrins from Urine
  • A10 Dipterex in Serum
  • A11 Diterpene Glycosides from Tea Extract
  • A12 Fatty Acids from Serum
  • A13 Homovanillic Acid from Plasma
  • A14 Isofl avones from Plant Extract
  • A15 Mixed Aromatic Amines from Urine and Plasma
  • A16 Nicotine and Cotinine from Plasma
  • A17 Nucleosides from Aqueous Solution
  • A18 Nucleosides from Plasma
  • A19 Nucleosides from Urine
  • A20 PEG 400 in Serum
  • A21 Persistent Organic Pollutants PAH, PCB and Pesticides in Blood
  • A22 Persistent Organic Pollutants PAH, PCB and Pesticides in Plasma
  • A23 Persistent Organic Pollutants PAH, PCB and Pesticides in Serum
  • A24 Pesticides and PCB in Fats
  • A25 Phenylanaline from Plasma
  • A26 Prostaglandins from Urine
  • A27 Prostaglandins from Whole Blood
  • A28 Salsoline from Plasma
  • A29 Steroid Acids from Serum
  • A30 Tryptophan from Plasma
  • A31 Vanillymandelic Acid from Plasma
  • A32 Xanthines (Caffeine) from Serum
  • A33 Xanthines (Theophylline) from Serum


ENVIRONMENTAL

  • E01 Carbamate Insecticide (Aldicarb) from Water
  • E02 PAH and PCB in Contaminated Soi
  • E03 Phenols in water
  • E04 Phthalate Esters in Water
  • E05 S-triazine Herbicide (Atrazine) in Soil


FOOD & BEVERAGE

  • F01 Afl atoxin B2 and M2 Metabolite Trace Analysis in Milk
  • F02 Chloroacetanilide Herbicides (Acetochlor and Metolachlor )in Contaminated Water
  • F03 F-2 Mycrotoxin Trace Analysis in Cereal
  • F04 Fatty Acid Methyl Esters (Long Chain) in Fermentation medium
  • F05 Omega 6 Fatty Acid in Malt Lipid Fractions
  • F06 Pigment Anthocyanidins in Wine
  • F07 S-triazine Herbicide (Atrazine) in Cereal
  • F08 Sulfonamide Trace Analysis in Meat


PHARMACEUTICALS

  • P01 Acetazolamide and Bumetanide in Urine
  • P02 Acetazolamide in plasma
  • P03 Amiodarone, Fendiline and Procainamide in Serum
  • P04 Amphetamine in Plasma
  • P05 Analgesics in Serum; Paracetamol and Tramadol
  • P06 Anesthetics in Serum; Benzocaine, Mepivacaine, Procaineand Lidocaine
  • P07 Antidepressants (Trycyclics) in Blood
  • P08 Antidepressants (Trycyclics) in Urine
  • P09 Atenolol in Plasma
  • P10 Barbiturates in Serum; Barbital, Amobarbital, Phenobarbital,Secobarbital
  • P11 Barbiturates in Urine; Barbital, Amobarbital, Phenobarbital,Secobarbital
  • P12 Carbamazepine, Phenobarbital and Primidone in Serum
  • P13 Chloramphenicol in Eye Drops
  • P14 Cimetidine in Plasma
  • P15 Cyclosporin in Blood
  • P16 Erythromycin and Clarithromycin in Urine
  • P17 Minor Tranquilizers (Benzodiazepines) in Urine or Serum
  • P18 Minor Tranquilizers (Diazepam and Lorazepam) in Hair
  • P19 Opiate Analgesics in Blood; Morphine and Codeine
  • P20 Propranolol in Serum
  • P21 Stobadin from Serum
  • P22 Vitamin A, D and E in Supplements
  • P23 Vitamin D3 and Metabolites in Serum

On-Line and Off-Line Application Of Micro-SPE (MEPS)

by Peter Dawes, Ern Dawes, Paul Wynne, SGE Analytical Science, and Dan DiFeo, SGE Inc.

Introduction Solid-phase extraction (SPE) has revolutionized sample preparation. Variations on the technique offer enhanced recovery, greater speciation and reduced solvent and sample consumption over other techniques. Micro-Extraction Packed Sorbent (MEPS) is the miniaturization of conventional SPE from milliliter to microliter bed volumes that allows SPE to be used with very small samples. The manipulation of the small volumes is achieved with a precision gas tight syringe. With a typical void volume of 7 L, the MEPS elution is compatible with GC and LC inlets making it ideal for integration into an automated sampling system for online SPE.

In most cases, MEPS allows the same level of sample concentration as is possible with off-line conventional SPE while providing opportunities for truly hybrid multi-dimensional methods. MEPS methods may be readily adapted from established SPE methods including those based on mixed mode or complex chemistries.

Figure 1. MEPS needle with integral 45-m C18 packed bed SPE device (void volume approximately 7 L). Click to enlarge Like SPE, MEPS is for use with liquid samples (either normal or reversed phase) and yields four fractions: the unretained, weakly bound, strongly bound and irreversibly bound. However, because MEPS is a double pass system (sample and solvent enter and exit from the bottom of the bed), the weakly bound fraction (commonly the interferences eliminated by washing) is less strongly bound. The irreversibly bound fraction affects MEPS and conventional SPE and is usually associated with sorbent wetting rather than sample purification and so the irreversible binding of matrix material from one sample does not preclude reuse of the device for a sample of the same type.

Like conventional SPE, the number of times the device can be reused is dependent on the sample matrix. For simple applications, MEPS devices

have been used successfully for 50 cycles.

Benefits of MEPS:

MEPS allows SPE methodology to be applied to small sample volumes. MEPS can be integrated into autosampler robotics and allows online use of SPE. MEPS can reduce sample and reagent consumption and waste disposal. Double pass flows can reduce the weakly bound fraction. MEPS is field portable for remote sampling with or without the use of automated equipment. MEPS is adaptable for other analytical techniques including immunoassay and off-line analysis by NMR, IR and other methods.

Plant Extracts Figure 2. Phytolacca octandra. The plant Phytolacca octandra (Figure 2) is one of the inkweeds or Pokes. In this application, the aerial portions of the plant were homogenized in acidified methanol, allowed to percolate for 12 hours, filtered and then extracted using a C2 MEPS cartridge. The MEPS was conditioned with methanol (30 L), water (30 L) and then 100 L of the plant extract was passed through the sorbent at 5mL/sec. The exhausted fraction was ejected at the same rate and the sorbent washed with 100mL of water. The sorbent was dried with air (3 80 L at 50 mL/sec) and eluted sequentially with hexane (10 L), dichloromethane (10 L) and methanol (10 L). The eluates were analyzed directly by GCMS on a BPX5 column (Figure 3).

The C2 MEPS method allowed the one step isolation of the FAME fraction (hexane) and the elimination of the highly polar sugar fraction (see Figure 3). Speciation of polar and non-polar analytes from a single sample digest was readily achieved without the need for off-line sample preparation.

Waste Water To demonstrate the usefulness of MEPS for dilute samples with a relatively simple matrix, a surrogate waste water sample was prepared from clear phenol free waste water spiked with either 25-ppb or 250-ppt phenols. The water was extracted with a C18 MEPS cartridge conditioned with methanol (30 L), water (30 L) and then 10 100 L of the water at 5 L/sec. The exhausted water was ejected at the same rate after each cycle; the sorbent was air dried (3 80 L at 50 L/sec). The analytes were eluted with methanol (10 L) and the fraction analyzed directly by GCMS on a BPX5 column. Chromatograms for the water samples are shown in Figures 4 and 5. Carryover was examined following the extraction of the 25-ppb sample by elution of a second and third portion of methanol without any intervening wash steps (Figure 5).

Figure 3. Effectiveness of the water wash in a GCMS analysis of Phytolacca octandra extract before and after fractionation on MEPS C2. Click to enlarge. Figure 4. Phenols in waste water at 250 ppt. Click to enlarge. Figure 5. Phenols in waste water at 25 ppb. Click to enlarge.

The C18 MEPS method allowed the one step isolation of phenols from water with good recovery, linearity and little carryover. The sorbent was reusable for the application for a large number of samples with no loss of performance after 10 analyses.

Toxicology Samples Papaver somniferum (opium poppy) is a feed contaminant that can result in positive drug tests for racing horses. We describe the extraction of a urine sample from an animal receiving contaminated feed as a demonstration of the off-line application of mixed mode C8/SCX MEPS for complex biological fluids.

Figure 6. Opiate metabolites in urine following MEPS of hydrolyzed urine. Click to enlarge. Here, 300 L of a diluted equine urine was hydrolyzed with b-glucuronidase or acid, filtered and extracted on a C8/SCX MEPS cartridge conditioned with methanol (30 L), potassium phosphate buffer (0.2 M, pH 6, 30 L) at a flow rate of 5 L/sec. The exhausted fraction was ejected at the same rate and the sorbent washed with 100-L phosphate buffer, 50-L acetic acid (1% v/v) and 100-L methanol. The sorbent was dried with air (3 80 L at 50 L/sec) and the sorbent eluted with 20-L dichloromethane-isopropanol-ammonia (49:49:2). The organic phase was evaporated under nitrogen and derivatized with 10 L of acetic anhydride-pyridine (1:2) at 80 C for 30 minutes before evaporation and reconstituted in 5 L of ethyl acetate. The extract was analyzed by GCMS on a BPX5 column (Figure 6).

The C8/SCX MEPS method allowed the microscale preparation of a small volume sample with comparable performance to conventional SPE techniques. Used off-line with derivatization and GCMS here, the sample was also suitable for online ESI-LCMSMS analysis by changing the elution solvent to methanol-ammonia (98:2) or methanol-trimethylamine (98:2).

For more information, contact Dan DiFeo, technical support, SGE Inc., at ddifeo@sge.com or by phone at 800-945-6154. Company’s Other Products Laboratory Equipment Rockaway NJ 07866

Videos of PAL


LEAP's PAL Application Videos on YouTube





 ↳ leaptec.com  ↳ leapwiki.com

LEAP Technologies provides fully automated sample processing workstation instrumentation solutions based on the LEAP CTC PAL X, Y, Z syringe only autosampler robot. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography. LEAP also offers process control systems and stand alone workstations. LEAP is an open, neutral resource for technical support of existing PAL. Keep your PAL in peak working order LEAP offers full support and service for the PAL platform with over 11 years of history with the PAL. LEAP can also write custom macros, cycles, and scheduling software for your specific applications. We're proud that LEAP is known as a productivity improver for laboratories including government labs in agencies such as FDA, USDA, EPA, FBI, Armed Forces, Homeland Security and others function more efficiently. Robotic sample prep equipment works well in government labs and challenges lab technicians to keep feeding those more samples. Analytical equipment such as chromatographs and mass spectrometers are expensive devices and also expensive to maintain. LEAP can help to keep them running and analyze samples around the clock. LEAP has most of its laboratory automation robots under GSAcontract and is eager to service and maintain everything it sells to keep your expensive lab equipment running around the clock. CTC Analytics Combi PAL and HTS PAL for Front-End Automation for LC and GC applications. LEAP offers customized on-site training. LEAP Is continuing to develop FAQ, Solutions, Tips & Techniques,a Learning Center with the LEAP Wiki for the benefit of all PAL Platform Robotic users. We will continue exceeding customer expectations for quality, cost effective and efficient results using the most current technology and expertise in our industry. Our aim is to be the ultimate knowledge base on everything related to the PAL. LEAP is known as the autosamplerguys being in business over twenty years. LEAP has extensive multi-vendor experience with integrating the PAL. HTX, HTS, HTC, HTX-xt, HTS-xt, HTC-xt, LC-Mini, Combi PAL,COMBI-xt PAL, and GC PAL models. We have the Quality, Training, Parts availability, Infrastructure, and Experience.LEAP is sometimes referred to as an Application Automation House who is continually adding new products and solutions to its Lab Automation line. As we do so we change the content of this web site so please check back.
Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions whether it is our standard applications or customized to your needs. The Combi-xt PAL autosampler is able to perform liquid and headspace analysis. Headspace options include Solid Phase Micro-extraction (SPME) and ITEX-2 as alternative to purge and trap. The HTS-xt PAL autosampler dramatically increases sample throughput by offering automated sample prep, Peltier cooled and heated sample storage using the cooled stack, Peltier cooled and heated sample storage using the cooled stack and low carryover due to the self-cleaning DLW (Dynamic Load and Wash) module and L-Mark syringes, peltier cooled valves and columns. LEAP provides automation that fits your needs and reduces errors, increases sample capacity, increases efficiency, reduces cost by supplying long term productivity improvements with user friendly software and flexible hardware. LEAP performs just in time or on time sample prep and inject. LEAP offers full support of integrated and non-integrated PAL software with both sales and support for Cruise Control, Cycle Composer and Cycle Composer Macro Editor, Agilent ChemStation CTC Control, Mass Hunter and EZChrom CTC integration, Xcalibur, Waters Masslynx or Analyst CTC Control software, Thermo (Finnegan) XCalibur, LEAP Shell, and Chronos software. The PAL is more than just an autosampler, it is a high-quality and efficient laboratory robot platform that can be customized by LEAP Technologies to meet your specific needs.

Contact LEAP

Contact LEAP

For additional information about LEAP and the PAL Platform, please contact LEAP Technologies.


The information and references contained in this web site are believed to be accurate and the most current information available.

Personal tools