Static Headspace Gas Chromatography – Mass Spectrometry(SHS-GC–MS) Method for Monitoring the Formation of ethyl methane sulfonate from ethanol and methane sulfonic acid
From LEAP
(New page: A Twin PAL [http://www.leapwiki.com/mediawiki/index.php?title=TwinPAL] is used for Static Headspace Gas Chromatography – Mass Spectrometry(SHS-GC–MS) Method for Monitoring the Formatio...) |
Current revision (20:42, 8 June 2010) (view source) |
||
(8 intermediate revisions not shown.) | |||
Line 1: | Line 1: | ||
- | A | + | A [http://www.leapwiki.com/mediawiki/index.php?title=TwinPAL Twin PAL] is used for '''Static Headspace Gas Chromatography – Mass Spectrometry(SHS-GC–MS) Method for Monitoring the Formation of ethyl methane sulfonate from ethanol and methane sulfonic acid'''<br> |
[[Image:HSTwin PAL.png|center]] | [[Image:HSTwin PAL.png|center]] | ||
'''TWIN PAL from LEAP System consists of:''' | '''TWIN PAL from LEAP System consists of:''' | ||
- | * 100 ul syringe | + | * 100 ul syringe adapter and syringe, 1 mL Head Space heated adapter and syringe, |
* Two (5) position 10 mL reagent holders | * Two (5) position 10 mL reagent holders | ||
* (6) position Agitator for 20 mL vials | * (6) position Agitator for 20 mL vials | ||
Line 9: | Line 9: | ||
* Chronos Mastersoftware with methods for sample prep and inject | * Chronos Mastersoftware with methods for sample prep and inject | ||
* Installation & Training | * Installation & Training | ||
- | ''Please note that the cooled tray uses an external chiller not supplied by LEAP'' | + | ''Please note that the cooled tray uses an external chiller not supplied by LEAP''<br> |
'''The available vial trays are:''' | '''The available vial trays are:''' | ||
* Tray 98 position temperature controlled for 2mL reaction vials | * Tray 98 position temperature controlled for 2mL reaction vials | ||
* Tray 32 position ambient tray for 20mL HS vials | * Tray 32 position ambient tray for 20mL HS vials | ||
- | * Tray 5 position for 10mL vials containing IS solution, derivatization reagent solution | + | * Tray 5 position for 10mL vials containing IS solution, [http://www.leapwiki.com/mediawiki/index.php?title=Category:Derivatization derivatization] reagent solution |
* Tray 5 position for 10mL vials containing 2 wash solutions and waste vials | * Tray 5 position for 10mL vials containing 2 wash solutions and waste vials | ||
- | The typical sample preparation sequence is as follows: | + | '''The typical sample preparation sequence is as follows:''' |
- | With 100ul liquid syringe | + | * With 100ul liquid syringe |
* Transfer 20uL reaction mixture from heated tray to 20 mL HS vial | * Transfer 20uL reaction mixture from heated tray to 20 mL HS vial | ||
* Wash syringe | * Wash syringe | ||
* Add 20uL IS solution from 10 mL reagent vial to 20 mL HS vial | * Add 20uL IS solution from 10 mL reagent vial to 20 mL HS vial | ||
* Wash syringe | * Wash syringe | ||
- | * Add 100uL | + | * Add 100uL [http://www.leapwiki.com/mediawiki/index.php?title=Category:Derivatization derivatization] solution from 10 mL reagent vial to 20 mL HS vial |
* Wash syringe | * Wash syringe | ||
- | With 1000ul headspace syringe | + | '''With 1000ul headspace syringe''' |
* Move 20 mL vial to heated agitator | * Move 20 mL vial to heated agitator | ||
* Heat 105 deg C and agitate at 600rpm for 15 minutes | * Heat 105 deg C and agitate at 600rpm for 15 minutes | ||
Line 30: | Line 30: | ||
---- | ---- | ||
[[Image:Chronos_Sample_Sequence.png|right]] | [[Image:Chronos_Sample_Sequence.png|right]] | ||
- | '''Info on Chronos software from LEAP Technologies''' | + | <br> |
- | Chronos Mastersoftware is an ideal solution for automating complex GC Sample Preparation steps on the CTC PAL. Software Integration with data systems ChemStation™ or Xcalibur™ assures seamless functionality with existing systems. Chronos not only exchanges information with other systems, but also controls the complete process and optimizes intelligent time-management resulting in significant improvement in system throughput.Single Sample List Control – including the flexible Twin PAL | + | '''Info on Chronos software from LEAP Technologies'''<br> |
+ | Chronos Mastersoftware is an ideal solution for automating complex GC Sample Preparation steps on the CTC PAL. Software Integration with data systems ChemStation™ or Xcalibur™ assures seamless functionality with existing systems. Chronos not only exchanges information with other systems, but also controls the complete process and optimizes intelligent time-management resulting in significant improvement in system throughput.Single Sample List Control – including the flexible Twin PAL<br><br><br> | ||
+ | [http://www.leapwiki.com/mediawiki/index.php?title=Category:Chronos Chronos Mastersoftware for organizing parallel processes for sample preparation] | ||
+ | <br><br> | ||
---- | ---- | ||
- | [[image:Download Icon.png|40px]] [[:media:Twin_Estimate_Rev_20091106. | + | |
+ | [[image:Download Icon.png|40px]] [[:media:Twin_Estimate_Rev_20091106.pdf| Download of this page]]<br> | ||
---- | ---- | ||
- | [[image:Accessories Icon.png|40px]] [http://leapwiki.com/mediawiki/index.php?title= | + | [[image:Accessories Icon.png|40px]] [http://www.leapwiki.com/mediawiki/index.php?title=GC_Headspace Headspace] |
+ | ---- | ||
+ | [[image:Accessories Icon.png|40px]] [http://www.leapwiki.com/mediawiki/index.php?title=TwinPAL Twin PAL] | ||
+ | |||
+ | [[image:Accessories Icon.png|40px]] [http://leapwiki.com/mediawiki/index.php?title=Cooling_and_Heating_Sample_Storage Cooling &Heating Sample Storage] | ||
[[image:Accessories Icon.png|40px]] [http://leapwiki.com/mediawiki/index.php?title=Mixing_of_Samples Mixing and Heating of Samples] | [[image:Accessories Icon.png|40px]] [http://leapwiki.com/mediawiki/index.php?title=Mixing_of_Samples Mixing and Heating of Samples] | ||
Line 44: | Line 52: | ||
[[image:Accessories Icon.png|40px]] [http://www.leapwiki.com/mediawiki/index.php?title=Category:Accessories Other Accessories for PAL Robots] | [[image:Accessories Icon.png|40px]] [http://www.leapwiki.com/mediawiki/index.php?title=Category:Accessories Other Accessories for PAL Robots] | ||
- | + | '''For other cost/time saving measures from LEAP consider:''' | |
- | |||
---- | ---- | ||
+ | === Videos of PAL === | ||
+ | <br> | ||
+ | [[Image:Movie Icon.png|40px]] | ||
+ | [http://www.youtube.com/user/LEAPTechnologies#g/u LEAP's PAL Application Videos on YouTube] | ||
+ | <br> | ||
{{logo}} | {{logo}} | ||
LEAP strives to find total solutions for analytical lab automation by automating analytical processes for small and large molecules in extracted liquids, solids, and recently in human, animal and plant tissues. We provide the precise robotics and efficient sample prep required by modern measurement techniques such as MALDITOF mass spectroscopy. Our newest specialty customization of CTC Analytics’ PAL features small workstations that can perform complex liquid handling tasks including HPLC-Purification, SPE, filtration, weighing, heating and stirring. They can be configured as stand alone units or integrated for “just in time” sample prep for LC-MS or GC-MS analysis. <br> | LEAP strives to find total solutions for analytical lab automation by automating analytical processes for small and large molecules in extracted liquids, solids, and recently in human, animal and plant tissues. We provide the precise robotics and efficient sample prep required by modern measurement techniques such as MALDITOF mass spectroscopy. Our newest specialty customization of CTC Analytics’ PAL features small workstations that can perform complex liquid handling tasks including HPLC-Purification, SPE, filtration, weighing, heating and stirring. They can be configured as stand alone units or integrated for “just in time” sample prep for LC-MS or GC-MS analysis. <br> |
Current revision
A Twin PAL is used for Static Headspace Gas Chromatography – Mass Spectrometry(SHS-GC–MS) Method for Monitoring the Formation of ethyl methane sulfonate from ethanol and methane sulfonic acid
TWIN PAL from LEAP System consists of:
- 100 ul syringe adapter and syringe, 1 mL Head Space heated adapter and syringe,
- Two (5) position 10 mL reagent holders
- (6) position Agitator for 20 mL vials
- Ambient tray for (32) 20 mL vials
- Cooled tray for (98) 2 mL vials
- Chronos Mastersoftware with methods for sample prep and inject
- Installation & Training
Please note that the cooled tray uses an external chiller not supplied by LEAP
The available vial trays are:
- Tray 98 position temperature controlled for 2mL reaction vials
- Tray 32 position ambient tray for 20mL HS vials
- Tray 5 position for 10mL vials containing IS solution, derivatization reagent solution
- Tray 5 position for 10mL vials containing 2 wash solutions and waste vials
The typical sample preparation sequence is as follows:
- With 100ul liquid syringe
- Transfer 20uL reaction mixture from heated tray to 20 mL HS vial
- Wash syringe
- Add 20uL IS solution from 10 mL reagent vial to 20 mL HS vial
- Wash syringe
- Add 100uL derivatization solution from 10 mL reagent vial to 20 mL HS vial
- Wash syringe
With 1000ul headspace syringe
- Move 20 mL vial to heated agitator
- Heat 105 deg C and agitate at 600rpm for 15 minutes
- Inject 1mL of Headspace into GC
- Move 20 mL vial back to ambient tray
Info on Chronos software from LEAP Technologies
Chronos Mastersoftware is an ideal solution for automating complex GC Sample Preparation steps on the CTC PAL. Software Integration with data systems ChemStation™ or Xcalibur™ assures seamless functionality with existing systems. Chronos not only exchanges information with other systems, but also controls the complete process and optimizes intelligent time-management resulting in significant improvement in system throughput.Single Sample List Control – including the flexible Twin PAL
Chronos Mastersoftware for organizing parallel processes for sample preparation
Cooling &Heating Sample Storage
Other Accessories for PAL Robots
For other cost/time saving measures from LEAP consider:
Videos of PAL
LEAP's PAL Application Videos on YouTube
LEAP strives to find total solutions for analytical lab automation by automating analytical processes for small and large molecules in extracted liquids, solids, and recently in human, animal and plant tissues. We provide the precise robotics and efficient sample prep required by modern measurement techniques such as MALDITOF mass spectroscopy. Our newest specialty customization of CTC Analytics’ PAL features small workstations that can perform complex liquid handling tasks including HPLC-Purification, SPE, filtration, weighing, heating and stirring. They can be configured as stand alone units or integrated for “just in time” sample prep for LC-MS or GC-MS analysis.
LEAP provides automated workstation instrumentation solutions based on the LEAP CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography.LEAP offers full support and service for the PAL platform in addition to being able to write custom macros, cycles, and scheduling to your applications. Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions.
Contact LEAP
Contact LEAP
For additional information about LEAP and the PAL Platform, please contact LEAP Technologies. |