Online Sample Cleanup using column switching with Multi-Valve system

From LEAP

(Difference between revisions)
Jump to: navigation, search
Line 6: Line 6:
----
----
-
[[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]]
+
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
=== Contact LEAP ===
=== Contact LEAP ===

Revision as of 21:44, 28 May 2009

A setup of four valves (1 injection valve and 1 TrioVS 3-valve system) is used to accomplish solid phase sample extraction by one (1) pre-column followed by chromatography on two (2) analytical columns. The goal is an increase in throughput by simultaneously executing sample preparation tasks and one of the two gradient runs. Gradient runs themselves do not occur in a staggered fashion.


Project Ref: H1099



Contact LEAP

For additional information about this technique please contact LEAP Technologies for detailed information