M-002 Dual Channel Serial Injection with front cut - Analyst with ICC-CE
From LEAP
Line 1: | Line 1: | ||
{{LSApp | {{LSApp | ||
|name = Dual Channel Serial Injection with Front Cut | |name = Dual Channel Serial Injection with Front Cut | ||
- | |image = | + | |image = 2_valve_on_top.png |
|type = '''SPECIAL''' | |type = '''SPECIAL''' | ||
|id = M-002 | |id = M-002 |
Revision as of 17:09, 25 March 2009
Overview
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run directly from within Analyst using a custom Cycle (requires ICC-CE). With the addition of a selector valve, the solvent front can be cut to waste.
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.
Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.
Contact LEAP
Keyword: Analyst Software, ICC-CE, PAL,TRIO VS, 10 port valve, selector valve, injector valve, gradient pump, regeneration pump,front cut, peaks of interest, GLP