M-002 Dual Channel Serial Injection with front cut - Analyst with ICC-CE
From LEAP
(Difference between revisions)
(→Overview) |
|||
Line 3: | Line 3: | ||
|image = M-Fab_Photo1.png | |image = M-Fab_Photo1.png | ||
|type = '''SPECIAL''' | |type = '''SPECIAL''' | ||
- | |id = M- | + | |id = M-002 |
|description = High throughput where one column is regenerating while the other is performing analytical separation. An additional valve allows solvent front to be cut to waste. | |description = High throughput where one column is regenerating while the other is performing analytical separation. An additional valve allows solvent front to be cut to waste. | ||
}} | }} |
Revision as of 01:39, 24 February 2009
Overview
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run directly from within Analyst using a custom Cycle (requires ICC-CE). With the addition of a selector valve, the solvent front can be cut to waste.
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.
Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.