- | === Cleaning the Leap Syringe with Fast Wash Station | + | === Cleaning the Leap Syringe with Fast Wash Station === |
This can clean the needle or help remove a bubble | This can clean the needle or help remove a bubble | ||
* Menu (F1) | * Menu (F1) | ||
Line 56: | Line 56: | ||
---- | ---- | ||
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]] | [[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]] | ||
+ | |||
=== Rinsing the Wash Station and Tubing with Solvent after refilling Solvent Reservoir === | === Rinsing the Wash Station and Tubing with Solvent after refilling Solvent Reservoir === | ||
* To replenish the Wash solvent, close the stopcock at the base of the Wash Bottle | * To replenish the Wash solvent, close the stopcock at the base of the Wash Bottle |
Revision as of 19:34, 2 July 2009
Training Videos | |
Application Type | |
Application ID | |
Description | |
Training Videos |
Contents |
Training Videos of PAL
Changing the Syringe of the PAL
Changing the Needle Sleeve of the PAL
Changing the Injection Depth at Valve of the PAL
Manual Injection of the PAL
Calibration of 96 well microtiter plate of the PAL
You will be leaving the LEAPWiki and going to the MicroNMR Website. To return just press the back button in your browser window.
The videos are FLASH videos and should be easily viewed within any standard web browser. Simply click on one of the link to play that video. If you are having trouble viewing the videos you may need to install the Flash Plug-in from the Adobe website (www.adobe.com).
There is AUDIO so make sure your speakers are on!
MicroNMR a partner of LEAP Technologies has put together some basic Training Videos for the PAL robot
A special thanks to MicroNMR for providing the videos on their website for all to access http://www.micronmr.com/training_videos.html
Leap Syringe Replacement
- Prime the replacement syringe manually. Use fresh (dry) solvent, and remove all bubbles.
- On the Leap Hand Held, hit Menu (F1), Chang Syr (F1) and wait for the Z-arm and syringe to position itself
- On the autosampler syringe arm, slide up the clear plastic cover.
- Near the top of the syringe, loosen the red knob by turning it to the right.
- At the midpoint of the syringe, rotate the small, black arm by 90 degrees to restrain the syringe body.
- Pinch together the pair of black needle guides that the needle slides through in normal use.
- With the needle guides pinched together, remove the syringe by gently tipping it forward and up.
- Put into place the new syringe, again pinching together the black needle guides.
- Be sure the syringe is against the mounting plate at its top and bottom.
- Rotate the small, black arm back to its original position.
- Tighten the red knob against the syringe plunger by turning the knob to the left.
- Slide closed the clear plastic cover.
- On the Leap Hand Held, hit the ESC key (repeatedly if needed) to return to the top of the menu.
- One at the top of the menu, hit F1 twice.
- Disable the syringe detection feature by following the on-screen directions.
- Enter the correct volume for the syringe presently in place.
Cleaning the Leap Syringe with Fast Wash Station
This can clean the needle or help remove a bubble
- Menu (F1)
- Clean Syr (F2) <enter>
- Select a Wash Station <enter>
- Enter the number of times the syringe should rinse <enter>
This option will move the LEAP syringe to the desired wash station and will rinse the full length of the syringe body. For new, or completely dry syringes, it is recommended to slightly pre-wet the syringe with solvent to displace some of the air.
Rinsing the Wash Station and Tubing with Solvent after refilling Solvent Reservoir
- To replenish the Wash solvent, close the stopcock at the base of the Wash Bottle
- Disconnect the tube at the black nut that goes into the white stopcock at the base of the bottle
- Rinse the bottle with fresh solvent and drain some solvent out through the stopcock
- Place the Wash bottle back into position, re-connect its tube, and open the stopcock
- Menu (F1)
- Utilities <enter>
- Wash Station <enter>
- Select desired wash station <enter>
- Press Act Valve (F2) to open the valve (located in the solenoid under the Wash Station)
- Rinse the tube until all the bubbles in it are flushed out (several seconds)
- Press Deact Valve (F2) to close the solenoid valve
- When F2 reads Act Valve, the solenoid is closed
- When reads Deact Valve, the solenoid is open
- Home
Switching the Valve
- Menu
- Utilities <enter>
- Injector <enter> (hit <enter> again to select default injector)
Pressing Act Valve (F2) puts the valve in the red light Off position = injection port connected to loop Pressing Deact Valve (F2) puts the valve in the red light On position = injection port not connected to loop
- Home
Changing Trays
- Menu (F1)
- Setup (F3) <enter>
- Objects <enter>
- Trays <enter>
- Select Desired Tray (usually Tray01, Tray02, Tray….) <enter>
- Scroll down to highlight ‘Tray Type’ <enter> and scroll through options
- VT54 – Vial Tray with 54 locations (blue)
- MT96 – Microplate Tray with 96 wells (short)
- DW96 – Deep Well tray with 96 wells (tall)
- Confirm/Adjust XYZ to coincide with Tray position 1 and well 1 (see detailed directions below)
- Be sure to save each coordinate
- Check the corners of the Tray
Setting Needle Penetration Depth for a Particular Tray
- Setting the Tray Holder XYZ position (performed at installation & described elsewhere) also enables the Needle Guide to find corner positions of a specified Tray (or well plate)
- This procedure guides a user to select a Tray, then adjust the needle penetration for a vial resting in that Tray. All of the XYZ positions for the Tray Holder, Wash Stations, Injector Port, and Waste should be calibrated prior to this procedure
- Menu (F1)
- Utilities <enter>
- Tray (select Tray – Tray## or Stack - Stk1-##) <enter>
- Needle Penetr <enter>
- Start by entering a Needle Penetr depth of 28.0 mm <enter>
- Movto 001 (F3), Needle Penetr <enter>, Dial depth downward till downward syringe motion halts, back up 0.1 mm, <enter>
- Movto 009 (F3), Needle Penetr <enter>, Dial depth downward till downward syringe motion halts, back up 0.1 mm, <enter>
- Movto 054 (F3), Needle Penetr <enter>, Dial depth downward till downward syringe motion halts, back up 0.1 mm, <enter>
- Other Tray Types will have different numbers for the corner positions
- Home
This information is not meant to replace on-site familiarization but provide a basic outline of the steps involved. Please refer to your PAL owners manual for detailed information.
LEAP offers installation and familiarization training with every new system. The twelve month warranty includes unlimited telephone support and module exchange programs. Extended warranties include annual preventative maintenance in addition to telephone support and module exchange programs.
For customers that want to write their own custom macros and injection cycles, LEAP offers software training at our facility in Carrboro, NC.
LEAP provides automated workstation instrumentation solutions based on the LEAP CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography.LEAP offers full support and service for the PAL platform in addition to being able to write custom macros, cycles, and scheduling to your applications. Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions.