XChange

From LEAP

(Difference between revisions)
Jump to: navigation, search
Line 32: Line 32:
Known limitations of XCHANGE® on the PAL-xt<br>
Known limitations of XCHANGE® on the PAL-xt<br>
-
1. XCHANGE® is only available for PAL-xt instruments.
+
1. XCHANGE® is only available for PAL-xt instruments <br>
-
2. Automatic Syringe recognition of PAL-xt must be turned on.
+
2. Automatic Syringe recognition of PAL-xt must be turned on <br>
-
3. There is only space for Wash Stations or Valves behind the Park Station. No Tray Holder or Stack can be placed behind the XCHANGE Park Station.
+
3. There is only space for Wash Stations or Valves behind the Park Station. No Tray Holder or Stack can be placed behind the XCHANGE Park Station. <br>
-
4. No magnetic vial transport possible with the actual design at this time.
+
4. No magnetic vial transport possible with the actual design at this time. <br>
-
6. Max Penetration Depth = 40mm  
+
6. Max Penetration Depth = 40mm <br>
-
7. Optional barcode reader excludes use of XCHANGE.  
+
7. Optional barcode reader excludes use of XCHANGE. <br>
-
9. XCHANGE is currently only available for the GC-xt, Combi-xt, HTS-xt instruments (80cm).
+
9. XCHANGE is currently only available for the GC-xt, Combi-xt, HTS-xt instruments (80cm). <br>
‎‎
‎‎
[[Image:XChange on 7890 A.png|400px|left|XChange by LEAP Technologies]] ‎
[[Image:XChange on 7890 A.png|400px|left|XChange by LEAP Technologies]] ‎

Revision as of 13:45, 23 March 2011

XChange
Application Type
  Applications
Application ID
  Syringe exchange
Description
  XCHANGE® Option for automated syringe exchange
 ↳ leaptec.com  ↳ leapwiki.com

XCHANGE® Option for automated syringe exchange

XChange by LEAP Technologies

New XCHANGE® Option which is fully implemented on the PAL-xt platforms

XCHANGE® is a coupling system easily integrated onto the PAL-xt platforms giving increased flexibility.

Automated syringe exchange enables new application possibilities with enhanced flexibility.
Main features are:
• Injection of small and large volumes in one sample sequence
• Adding of internal standards without cross contamination
• Preventative exchange of syringe when using harmful media
• Choice of 9 different syringe volumes
• Serial dilution

Automated syringe exchange enables new application possibilities with enhanced flexibility.
The XCHANGE® coupling system secures and locks the syringe in place. This is achieved by a mating mechanical piece that captures the syringe and a magnet that automatically aligns – eliminating the need for springs, screws, or alignment gauges. Once engaged the syringe is locked into position requiring zero holding force.

XCHANGE ® syringes (10µl, 25ul, 100µl, 250ul, 500ul, 1ml, 2.5ml and 5ml)
(requires PAL Firmware 4.2.x)

Known limitations of XCHANGE® on the PAL-xt

1. XCHANGE® is only available for PAL-xt instruments
2. Automatic Syringe recognition of PAL-xt must be turned on
3. There is only space for Wash Stations or Valves behind the Park Station. No Tray Holder or Stack can be placed behind the XCHANGE Park Station.
4. No magnetic vial transport possible with the actual design at this time.
6. Max Penetration Depth = 40mm
7. Optional barcode reader excludes use of XCHANGE.
9. XCHANGE is currently only available for the GC-xt, Combi-xt, HTS-xt instruments (80cm).
‎‎

XChange by LEAP Technologies








Contact LEAP

For additional information about LEAP and the PAL Platform, please contact LEAP Technologies.

Keywords

SGE, CTC, XCHANGE® coupling system secures and locks the syringe in place, Pittcon 2011, LabAutomation 2011,XCHANGE ® syringes (10µl, 100µl, 1ml and 5ml,requires PAL Firmware 4.2.x,PAL-xt platforms

Personal tools