Dual Channel Sample Cleanup

From LEAP

(Difference between revisions)
Jump to: navigation, search
Line 21: Line 21:
* Summary of LC Multi Valve Applications [http://www.leapwiki.com/mediawiki/index.php?title=Summary_of_LC_Multi_Valve_Applications]
* Summary of LC Multi Valve Applications [http://www.leapwiki.com/mediawiki/index.php?title=Summary_of_LC_Multi_Valve_Applications]
* LEAP Shell Software and application for custom injections and scheduling [http://www.leaptec.com/software/leap-shell-3.php]
* LEAP Shell Software and application for custom injections and scheduling [http://www.leaptec.com/software/leap-shell-3.php]
-
-----
+
----
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
Line 28: Line 28:
For additional information about this technique please contact LEAP Technologies for detailed information
For additional information about this technique please contact LEAP Technologies for detailed information
-
* [http://www.leaptec.com/contact-us.php Contact the LEAP office]
+
[[image:Contact_Info.png‎|40px]] [http://www.leaptec.com/contact-us.php Contact the LEAP office]
-
 
+
[[Category:Example Solutions]]
-
<br><br>
+
[[Category:LEAP Shell]]
[[Category:LEAP Shell]]

Revision as of 13:56, 9 June 2009

General Description of process:

Samples are injected and pass through a guard column which is in-line with the analytical column. At a specified time, the guard column switches out of line and is regenerated using a wash pump configured to do a step gradient while the sample peaks are acquired.

To increase throughput, samples are injected ahead of the acquisition time, so that pre-filtration is occurring concurrently with the previous samples acquisition. Each line of the sample list processes a pair of samples which are staggered in this way.

4 pumps are required: - 2 wash pumps (Agilent 1200) pre-programmed to perform the step gradient and 2 isocratic analytical pumps running continuously one for each column.


Image:Dual_Channel_Sample_Filtration_PlumbDiag.png


Project Ref: H1073b


For other cost/time saving measures consider:

  • LEAP offers multiple choices of Valve Self Wash Stations to aid in cleaning the valve [1]
  • “Look Ahead”, sample prep before injection (M-004) using LEAP Shell Software [2]
  • Dual Channel Parallel Injection (M-005) using LEAP Shell software [3]