Dual Channel Sample Cleanup

From LEAP

(Difference between revisions)
Jump to: navigation, search
Line 16: Line 16:
-----
-----
-
[[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]]
+
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
=== Contact LEAP ===
=== Contact LEAP ===

Revision as of 21:41, 28 May 2009

General Description of process:

Samples are injected and pass through a guard column which is in-line with the analytical column. At a specified time, the guard column switches out of line and is regenerated using a wash pump configured to do a step gradient while the sample peaks are acquired.

To increase throughput, samples are injected ahead of the acquisition time, so that pre-filtration is occurring concurrently with the previous samples acquisition. Each line of the sample list processes a pair of samples which are staggered in this way.

4 pumps are required: - 2 wash pumps (Agilent 1200) pre-programmed to perform the step gradient and 2 isocratic analytical pumps running continuously one for each column.


Image:Dual_Channel_Sample_Filtration_PlumbDiag.png


Project Ref: H1073b



Contact LEAP

For additional information about this technique please contact LEAP Technologies for detailed information




Personal tools