Valve Self Washing II
From LEAP
(Difference between revisions)
(→Contact LEAP) |
|||
Line 6: | Line 6: | ||
|description = Accessory Showcase | |description = Accessory Showcase | ||
}} | }} | ||
+ | |||
+ | [[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]] | ||
=== Main Features === | === Main Features === | ||
Line 51: | Line 53: | ||
*[[:Image:Valve_Self_Wash_II.pdf | PDF of this page formatted for printing]] | *[[:Image:Valve_Self_Wash_II.pdf | PDF of this page formatted for printing]] | ||
- | == Contact LEAP == | + | ----- |
+ | [[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]] | ||
+ | |||
+ | === Contact LEAP === | ||
For additional information about this technique please contact LEAP Technologies for detailed information | For additional information about this technique please contact LEAP Technologies for detailed information | ||
Line 58: | Line 63: | ||
- | [[Category: | + | <br><br> |
- | [[ | + | [[Category:Accessories]] |
+ | [[Valve Washing]] |
Revision as of 01:21, 29 April 2009
Valve Self-Washing System – II | |
Application Type | |
Washing LC Valve | |
Application ID | |
Valve Self-Washing System – II | |
Description | |
Accessory Showcase |
Contents |
Main Features
- Increased Throughput by simultaneous washing of injection syringe and injection valve
- 4 Wash Solvents can be used to wash the valve
- 9 Program Steps to wash with the 4 wash solvents in any order
- Software Interface SDS Launchpad
- Save/Load Methods
- Easy Verification of Downloaded Method via the “Dashboard” feature in SDS Launchpad
- Contact Closure Triggered
- Runs stored wash routine
- Washes independent of injection syringe
- Can increase syringe life
- Compatible with all PAL Drivers
Performance Specifications
- Maximum flowrate 6mL/min in increments of 0.1mL/min
- Flowrate accuracy ± 10%
- Maximum dispensed volume 999mL
- Volumetric accuracy ± 5%
- Maximum pressure 50PSIG
- Typical wash time for 2 solvents (1mL each) @ 6mL/min is 45 seconds
Requirements/Limitations
- Through-loop washes only for injection loop volumes > 20uL
- Lower dispense speeds needed for more viscous solvents, e.g., i-Propanol @ 3-4mL/min
- Use of Polypropylene compatible solvents
Screenshots
Other Reference material
PRINT Function for Page