Dual Channel Sample Cleanup
From LEAP
Line 17: | Line 17: | ||
=== Contact LEAP === | === Contact LEAP === | ||
- | * [http://www.leaptec.com/ | + | * [http://www.leaptec.com/contact-us.php Contact the LEAP office] |
Revision as of 14:18, 21 April 2009
General Description of process:
Samples are injected and pass through a guard column which is in-line with the analytical column. At a specified time, the guard column switches out of line and is regenerated using a wash pump configured to do a step gradient while the sample peaks are acquired.
To increase throughput, samples are injected ahead of the acquisition time, so that pre-filtration is occurring concurrently with the previous samples acquisition. Each line of the sample list processes a pair of samples which are staggered in this way (see time chart at the end of this document).
4 pumps are required: - 2 wash pumps (Agilent 1100) pre-programmed to perform the step gradient and 2 isocratic analytical pumps running continuously one for each column.
Project Ref: H1073b